Mobile Assemble Lab Function Division

- May 13, 2021-

Mobile Assemble Lab Function Division

1. Air conditioning purification and functional zoning are designed and

 manufactured in strict accordance with P3 (BSL-3) mobile laboratory standards,

 with three core work areas: reagent preparation area, sample preparation area

 and product amplification analysis area, all with independent buffer room; Two

 equipment rooms: clean equipment room (including gas, water purification, fresh

 air conditioning and UPS power supply), sewage equipment room (including

 sewage, exhaust); Also optional to add a separate wash room.

 

2. Built-in positive pressure clean corridor, easy to ensure the cleanliness of the

 cabin and observe the situation of each area.

 

3. The core working room is designed with negative pressure and certain pressure

 gradient. The air distribution is in accordance with the reagent preparation area

 -- sample preparation area -- product amplification and analysis area, and the air

 flow is in the form of one-way flow, full fresh air and full exhaust air.

 

4. Each core workshop is equipped with transfer Windows for goods transfer. The

 transfer Windows and doors have interlocking function and are equipped with UV

 lamp for disinfection.

5. All core workrooms are equipped with LED clean lights and UV lights, which are

 energy saving and easy to maintain, and can effectively prevent the breeding of

 bacteria. The room illumination meets the standard requirements; The ultraviolet

 lamp is equipped with delay function to protect the safety of the experimenter.

 

6. The door of the workshop equipped with differential gauge, can real-time

 display the dynamic pressure inside the room