Mobile Assemble Lab Function Division
1. Air conditioning purification and functional zoning are designed and
manufactured in strict accordance with P3 (BSL-3) mobile laboratory standards,
with three core work areas: reagent preparation area, sample preparation area
and product amplification analysis area, all with independent buffer room; Two
equipment rooms: clean equipment room (including gas, water purification, fresh
air conditioning and UPS power supply), sewage equipment room (including
sewage, exhaust); Also optional to add a separate wash room.
2. Built-in positive pressure clean corridor, easy to ensure the cleanliness of the
cabin and observe the situation of each area.
3. The core working room is designed with negative pressure and certain pressure
gradient. The air distribution is in accordance with the reagent preparation area
-- sample preparation area -- product amplification and analysis area, and the air
flow is in the form of one-way flow, full fresh air and full exhaust air.
4. Each core workshop is equipped with transfer Windows for goods transfer. The
transfer Windows and doors have interlocking function and are equipped with UV
lamp for disinfection.
5. All core workrooms are equipped with LED clean lights and UV lights, which are
energy saving and easy to maintain, and can effectively prevent the breeding of
bacteria. The room illumination meets the standard requirements; The ultraviolet
lamp is equipped with delay function to protect the safety of the experimenter.
6. The door of the workshop equipped with differential gauge, can real-time
display the dynamic pressure inside the room